The AUTO-PS-1 is a 30 L class semi-automatic magnetic bead preparation system. It scales the workflow you already run on a benchtop magnetic rack — load, bind, separate, wash and elute — up to 1–30 L per batch, inside a single floor-standing cabinet with a full glass door.
Scale-up is not simply a bigger vessel. Above roughly 10 L the liquid layer the field must act through is tens of times thicker, and bead settling, impeller shear and clarification time all behave differently. The AUTO-PS-1 addresses this with a ring magnet array encircling the vessel from outside, driving beads to the wall; the magnet assembly is raised and lowered by handwheel, so the operator keeps direct control of approach speed and dwell position. Agitation is top-mounted mechanical stirring, not a magnetic bar, so high-solids bead slurries do not slip.
"Semi-automatic" means reagents are added manually and separation, stirring and timing are operator-driven. There is no HMI and no software. It suits teams still mapping out their scale-up process, or sites without a computerised-system validation requirement yet.
The cabinet is organised in three zones, left to right:
| Position | Description |
|---|---|
| ① Reaction vessel | 30 L vessel, 316L lining, thermal/light-shielding enclosure (sight glass optional); top-mounted stirrer enters through the centre of the head, impeller above the bottom |
| ② Reagent positions | Two internal shelves, six 1 L positions (2 × 3), manually filled |
| ③ Ring magnet array | Black ring magnet assembly encircling the separation vessel, handwheel lift; beads collect on the vessel wall during separation |
| ④ Manual valve block | Central stainless valve block at the base; handwheels route feed, transfer and drain |
An annotated configuration drawing follows this section in the body copy. Left vessel = reaction vessel; right vessel with the black magnet ring = separation vessel; centre = reagent shelving.
| Parameter | Specification |
|---|---|
| Model | AUTO-PS-1 (Semi-automatic) |
| Batch volume | 1 – 30 L / batch |
| Separation method | Ring magnet array, handwheel lift |
| Separation time | ≤ 15 min (30 L aqueous) |
| Stirring | Top-mounted mechanical stirring, 0 – 200 rpm |
| Reaction vessel | 316L vessel lining + thermal/shielding enclosure; sight glass optional |
| Wetted materials | 316L SS / PTFE / Borosilicate glass 3.3 |
| Reagent positions | 6 positions (2 tiers × 3, 1 L bottles, manual filling) |
| Control | Manual / mechanical operation, no control panel |
| Dimensions | Approx. 1800 × 750 × 1350 mm (excl. external tubing) |
| Power | AC 220 V, 1.5 kW |
Specifications are subject to the technical documentation shipped with the unit. Items marked "optional" are not part of the standard configuration. We recommend a small-scale trial before scale-up; separation time can be verified on your own feed material.
| Step | Action |
|---|---|
| ① Load | Beads and feed are charged to the reaction vessel; six reagent bottles are loaded by hand |
| ② Bind | Top-mounted stirring at 0–200 rpm for the incubation step |
| ③ Separate | Lower the ring magnet assembly by handwheel; beads collect on the wall and the supernatant clears (≤ 15 min for 30 L aqueous) |
| ④ Wash and elute | Drain the supernatant, change over to wash or elution buffer by hand, repeat stir/separate, then collect beads or eluate |